MacView IPR

MacView IPR

The MACView® IPR is a universal measuring system for gasses, vapours and solid (dust) particles. A extra unique feature is that the system is controlled and the visualisation is processed by a integrated webserver.

With this system it is possible to follow facilities with a internet connection on one location, more office buildings, factory buildings or worldly spread production plants. The advantages are more then one. A Quality Assurance departement can check the performance of the air cleaning facility in a glance, or a Quality Control departement can monitor continuously the dust load in a packaging compartement. But also international operating companies which want to follow on a distance in different countries the performances of their production plants in real time, it is all possible.

The MACView® IPR is build modular. In theory one can add as much sensors as available internet adresses. The system is build in a 19” housing. Every 19” housing can position 8 IPR cards with each 2 sensors. This means that with one 19” housing 16 sensors can be monitored at the same time. When the 19” housing is full, it is easy to add another housing in the rack.

Number of slots8 slots can position 8 IPR cards in a 19” housing, housing is 3U high
Number of channels2 channels on one IPR card
Number of sensorsMore than 80 sensors available for gasses, vapours and fine dust particles, 3 wire 4-20mA-Combustibles - Toxic Gases - Oxidizing Gases - Organic Solvents - CFC's (HCFC's en HFC's) - Indoor Pollutants - Cooking Vapors – OxygenEvery sensor measures the following values of a quantity : concentration, relative humidity and temperature
Type of sensors206402: MACView® Wall-mount sensor 4-20mA output, gas, RH and T206502: MACView® Wall-mount EeX sensor 4-20mA output, gas, RH and T206304: MACView® Particles Advanced sensor 4-20mA output, fine dust particles, RH and T206305: MACView® Particles Advanced press fitting sensor 4-20mA output, fine dust particles, RH and T, suitable for ventilation channels
Housing materialHigh quality aluminium with anodising layer
StandardsNEN-EN-IEC 61000-6-1 t/m 4, CE, Ethernet IEEE 802.3, TCP/IP, HTML en SMTP
Signals/warningsppm of mg/m³ and hysteresis adjustable on relais (programmable for every function)Analogue output 4-20mA (programmable for every function)E-mail adjustableSeveral status LED’s : power, failure, RX, TX, Collision, relais 1, relais 2
Acceptance of warningsWith accept button on front side or through browser
Communication protocolTCP/IP over IEEE 802.3 Ethernet, (UTP connection with 8 wires)
RegistrationInternal database with date and time clock
Sensor inputs4-20mA suitable for long range with transmission protocol
Sensor connections3-wire : power, 0V, signal4-wire : power, 0V, signal, signal 0V (reference)
Inputs1 digital input (potential free)
Outputs2 analogue outputs (0-10V, 0-20mA or 4-20mA, adjustable throug the software) mA outputs are up to a load of 400Ohm at 12V2 digital outputs (potential free relais 230V, 1A) for concentration warning1 digital output (potential free relais 230V, 1A) for internal failure warning
Service connectionRS232 interface
PowerconnectionIn: 230 VAC
Out: 12V, 150W for external sensors
Operating temperature sensors-30 / +80 degree Celcius, relative humidity 5 till 95%, not condensed
Operating temperatureMACView® IPR-10 / +50 degree Celcius, relative humidity 5 till 95%, not condensed
SoftwareIntegrated in a autonomous webserverFor more IPR systems a control tool is available. This tool is not neccesary when using a small number of MACView® IPR systems
User interfaceHTML pages in a browser and E-mail
DimensionsMACView® IPRRack 3U, 19” housing W 482 x H 132 x D 266 mm
MountingIn a cabinet frame rack suitable for 19” housings